SOP FOR CALIBRATION OF FRIABILITY TEST APPARATUS

 FRIABILITY OF TABS MEANS—

  • TO CHECK SURFACE ABRATION OF TABS
  • CRUSHING STRENGTH TEST—MAY NOT BE BEST MEASURED OF POTENTIAL TABS, BEHAVIOUR DURING HANLING & PACKING.                          

CALIBRATION OF FRIABILITY TEST APPARATUS

ITEMS NO –                                                  MODEL NO –EF-1W

MAKE—ELECTROLAB                             LOCATION—

INS ID NO –                 FREQUENCY OF CALIBRATION— 3 MONTHS

CALIBRATION REPORT NO—

ACCEPTANCE CRITERIA = ± 1

 

CALIBRATION TIME SETTING  KNOB ON: 100 REVOLUTION IN 4 MIN , 25RPM

 

SR.NO                              OBS.REVULATION AFTER 4 MIN                    TOLERANCE     

1                                                   100                                                       99 to 101

2                                                   100                                                      99 to 101  

3                                                    100                                                     99 to 101

MEAN                                          100                                                       99 to 101

 

REMARKS—COMPLY/DOES NOT COMPLY

THE INS IS CALIBRATED AS PER THE WRITTEN PROCEDURE & COMPLIES/DOES NOT COMPLIES WITH THE ACCEPTANCE CRITERIA.

CALIBRATION SUMMARY –

CLEANLINESS OF THE INSTRUMENT:–

OBSERVATION: CLEAN / NOT CLEAN

CORRECTIVE ACTION IN THE CASE OF NON-COMPLIES

ANY OTHER MAINTENANCE:

OBSERVATION : REQUIRED /NOT REQUIRED

CORRECTIVE ACTION IN THE CASE OF NON-COMPLIES.