OBJECTIVE:-
TOASSURE THE WEIGHING BALABNCES BEING IN THE MANUFACTURING ACTIVITIES CONFIRM TO THE SPECIFIED ACCEPTANCE CRITERIA.
CALIBRATION SHALLBE PERFORMED AS:–
- ASPER SPECIFIED FEQUENCY.
- FOR NEW BALANCE DURING INSTALLATION.
- AFTER ANY MAJOR REPAIR OR COMPONENT REPLACEMENT.
PROCEDURE:–
- ENSURE THAT THE WEIGHING IS PROPERLY CLEANED AND LEBELLED.
- OPERATE THE BALANCE ASPER RESPECTIVE SOP.
- FOR CALIBRATION BALANCES ARE TO BE SUBJECTED FOR THE FOLLOWING TESTS.
- LINEARITY TEST
- ECCENTRICITY TEST.
- MEASUREMENT OF UNCERTAINTY (RANDOM PLUS SYSTEMATIC ERROR).
LINEARITY TEST: —
DIVIDE THE FULL SPAN OF BALANCE INTO 5 PARTS AND CONSIDER THE 5 PARTS AS STANDARD WEIGHTS FOR CALIBRATION.
1ST PART – ZERO
2ND PART – 25% OF FULL RANGE.
3RD PART – 50%OF FULL RANGE.
4TH PART – 75% OF FULL RANGE.
5TH PART – 95% OF FULL RANGE.
- PLACE THE WEIGHTS ON THE BALANCE ONE BY ONE IN INCREASING ORDER FOR THE 5 DIFFERENT STANDAR RANGES.
- REMOVE THE WEIGHTS IN THE REVERSE ORDER FROM THE BALANCE.
- CHECK THE DIFFERENCE B/W THE RESPECTIVE WEIGHTS OBSERVED IN BOTH THE ORDER FOR ANY INDIDUAL SET WEIGHT LIMIT.
ECCENTRICITY TEST :–
- PERFORM THE TEST USING ONE QURTER WEIGHT OF THE CAPACITY OF THE BALANCE.
- PLACE THE SELECTED WEIGHTS ON THE PLATEFORM AT 5 LOCATIONS AS ILLUSTRATED.
MEASUREMENT OF UNCERTAINTY (RANDOM PLUS SYSTEMIC ERROR):–
- THE WEIGHTS TO BE USED FOR TEST SHALL BE THE SINGLE MAXIMUM WEIGHT AVAILABLE NEAREST TO THE CAPACITY OF THE BALANCE.
- PLACE THE WEIGHT ON PAN/PLATFORM OF THE BALANCE.
- RECOPRD THE WEIGHT DISPLAYED.
- REMOVE THE WEIGHT AND AGAIN PLACE THE WEIGHT ON THE PAN AND RECORD THE DISPLAY ONCE AGAIN.
- REPEAT THE ABOVE OPERATION OF TEN READINGS AND RECORD THE OBSERVATIONS.
- CALCULATE THE STANDARD DEVIATION OF TEN READINGS. CALCULATES MEASUREMENT UNCERTAINTYAS GIVEN BELOW—
- MEASUREMENT UNCERTAINTY = RANDOM ERROR +SYSTEMIC ERROR X 100/ACTUAL WEIGHT
- RANDOM ERROR = 3 X STANDARD DEVIATION (n = 10)
- SYSTEMIC ERROR = OBSERVED WEIGHT (AVERAGE) – ACTUAL WEIGHT.
ACCEPTANCE CRITERIA –
LINEARITY TEST—THE DIFFERENCE IN DISPLAYED WEIGHT FOR ANY SET OF WEIGHTS DURING INCREASING & DECREASING ORDER SHALL BE NOT THAN PLUS –MINUS 2 LEAST COUNTS OF THE BALANCE.
ECCENTRICITY TEST –
THE DISPLAYED WEIGHTS SHALL NOT BE MORE THAN PLUS –MINUS LEAST COUNTS FROM THE STANDARD WEIGHT.
MEASUREMENT OF UNCERTAINITY (RANDOM PLUS SYSTEMIC ERROR)—
- THE STANDARD DEVIATION OF THE TEN CONSECUTIVE READINGS SHALL NOT BE MORE THAN ONE LEAST COUNT.
- THE MEASUREMENT UNCERTAINITY SHOULD NOT BE MORE THAN PLUS –MINUS 0.1%.